Detection of Bronchial and Arterial Obstruction by Continuous Gas Analysis from Individual Lobes and Segments of the Lung

P. Hugh‐Jones, John B. West

Thorax · 1960 · 29 citations · 7 references

DOIFull text

Open access

Concepts

Abstract

A mass spectrometer was designed specially for clinical respiratory investigation (Fowler, 1957) and it has been in use for general tests of lung function for three years. It performs continuous automatic gas analysis, and has a gas-sampling tube of only about I mm. external diameter at its tip. This fine sampling tube, which is placed at the patient's lips during general tests of lung function, has now been adapted so that it can be inserted down a bronchoscope into any lobar or segmental orifice which can be seen directly. From the results of bronchoscopic experiments on anaesthetized dogs it has been found that characteristic patterns of gas concentration changes are observed from regional bronchi during the breathing cycle and that these can be related to the ventilation, blood flow, and mechanical properties of the relevant lobe or segment. In this paper we shall first summarize these results from animal experiments and then present some examples of these characteristic patterns seen during routine diagnostic bronchoscopy in man and discuss their significance.

References