IEEE Transactions on Biomedical Circuits and Systems · 2010 · 30 citations · 19 references
EngineeringBiomedical EngineeringElectrical EngineeringCircuit TheoryEnergy SystemsPower ElectronicsEnergy ConversionHigh-frequency Power ConversionMaximum Power TransferPower System DesignResonant ConvertersSoft-switching TechniquesDesign Of Energy Conversion SystemsEnergy TransferMaximum Power OutputResonant Energy TransferResonant FrequencyResonant Frequency TuningResonant Power TransferTurning Point Analysis
Transcutaneous energy transfer (TET) enables the transfer of power across the skin without direct electrical connection. It is a mechanism for powering implantable devices for the lifetime of a patient. For maximum power transfer, it is essential that TET systems be resonant on both the primary and secondary sides, which requires considerable design effort. Consequently, a strong need exists for an efficient method to aid the design process. This paper presents an analytical technique appropriate to analyze complex TET systems. The system's steady-state solution in closed form with sufficient accuracy is obtained by employing the proposed equivalent small parameter method. It is shown that power-transfer capability can be correctly predicted without tedious iterative simulations or practical measurements. Furthermore, for TET systems utilizing a current-fed push-pull soft switching resonant converter, it is found that the maximum energy transfer does not occur when the primary and secondary resonant tanks are "tuned" to the nominal resonant frequency. An optimal turning point exists, corresponding to the system's maximum power-transfer capability when optimal tuning capacitors are applied.
19