Harbour porpoise (<i>Phocoena phocoena</i>) static acoustic monitoring: laboratory detection thresholds of T-PODs are reflected in field sensitivity

Line Anker Kyhn, Jakob Tougaard, Jonas Teilmann, Magnus Wahlberg, Poul Boel Jørgensen, Nikolaj I. Bech

Journal of the Marine Biological Association of the United Kingdom · 2008 · 45 citations · 15 references

Concepts

Abstract

The T-POD (Timing POrpoise Detector) is a self-contained acoustic data logger used for detecting and monitoring the presence of echolocation clicks of small cetaceans. It has become a standard tool in environmental impact assessments and monitoring programmes. Yet, little is known about the variability in sensitivity and detection range of T-PODs. In this study the field performance of ten v3 T-PODs was compared to detection thresholds measured in a tank. The T-POD thresholds ranged from 123 to 132 dB re 1μPa (pp). The detection thresholds of the ten individual T-PODs were different and the differences increased over time. The more sensitive a T-POD was in the laboratory (i.e. the lower the threshold) the more clicks were recorded by it in the field. Threshold correlated differently to the five analysed T-POD parameters (encounters, encounter duration, waiting time, porpoise positive minutes, clicks per porpoise positive minute). This study demonstrates that individual threshold calibrations of T-PODs are necessary to obtain comparable results when monitoring odontocetes with this tool. Regression equations for relationships between T-POD detection thresholds and study parameters obtained during field trials may allow comparisons of T-PODs with different detection thresholds.

References

15