Concepedia

Publication | Open Access

Imaging Capabilities of the Inveon SPECT System Using Single-and Multipinhole Collimators

23

Citations

18

References

2013

Year

Abstract

The acquisition parameters, such as the collimator set and the radius of rotation, offer a wide range of possibilities to apply to a large number of biologic studies. However, special care must be taken because this increase in sensitivity can be offset by image degradation, such as image artifacts caused by projection overlap and statistical noise due to a higher number of iterations required for convergence. 3DMAP allowed better noise reduction while maintaining relatively constant recovery coefficients, as compared with other reconstruction strategies.

References

YearCitations