Asian Spine Journal · 2011 · 18 citations · 19 references
MedicineOrthopedic SurgerySurgeryOrthopedicsPediatric OrthopaedicsSpine SurgeryLumbar Spine SurgeryPediatric ScoliosisPediatric Spine SurgerySpinal InstrumentationThoracic Spine SurgeryElectromyography InterpretationPediatric ElectromyographyPediatric Scoliosis SurgeryRadiographic Evaluation Of ScoliosisThoracic Pedicle Screw FixationThresholdsTriggered ActivityTriggered Electromyography
STUDY DESIGN: We performed a prospective study to evaluate the reliability of using triggered electromyography (EMG) for predicting pedicle wall breakthrough during the placement of pedicle screw in adolescent idiopathic scoliosis surgery. PURPOSE: We wanted to correlate pedicle wall breakthrough with the triggered EMG threshold of stimulation and the postoperative computed tomography (CT) findings. OVERVIEW OF LITERATURE: Pedicle wall breakthrough has been reported to be difficult to evaluate by radiographs. Triggered EMG had been found to be a more sensitive test to detect this breakthrough. METHODS: Seven patients who underwent the insertion of 103 pedicle screws were evaluated. The triggered EMG activity was recorded from several muscles depending on the level of screw placement. The postoperative CT scans were read by a spine surgeon who was a senior fellow in orthopedics, and a musculoskeletal radiologist. RESULTS: The mean age at surgery was 12.6 years (range, 11 to 17 years). The preoperative mean Cobb angle was 54.7° (range, 45 to 65°). There were 80 thoracic screws and 23 lumbar screws. All the screws had stimulation thresholds of ≥ 6 mA, except 3 screws with the stimulation threshold of < 6 mA. Ten screws (9.7%) showed violation of the pedicle wall on the postoperative CT scans. Five screws penetrated medially and another five penetrated laterally. No postoperative neurologic complications were noted in any of the seven patients. CONCLUSIONS: Measuring the stimulation threshold of triggered EMG helps to assess the pedicle screw placement. Pedicle screws that had stimulation threshold of ≥ 6 mA were safe, with 90.3% reliability, as was assessed on the postoperative CT scans.
19
Analysis of the Morphometric Characteristics of the Thoracic and Lumbar Pedicles
Michael R. Zindrick, Leon L. Wiltse, ALBERT DOORNIK et al. · Spine · 1987 · 540 citations
Pedicle Screw Instrumentation of the Thoracic Spine in Idiopathic Scoliosis
Ulf Liljenqvist, H. Halm, Thomas M. Link · Spine · 1997 · 486 citations
Placement of pedicle screws in the thoracic spine. Part I
Alexander R. Vaccaro, Steven J. Rizzolo, Thomas J. Allardyce et al. · Journal of Bone and Joint Surgery · 1995 · 349 citations
Accuracy of Computer-Assisted Pedicle Screw Placement