1986 · 23 citations · 10 references
Natural SciencesBiomedical EngineeringDiagnostic ImagingUltrasonographyUltrasoundAcoustic ImagingDiagnostic UltrasonographyImagingUltrasonicsUltrasound ImagingClutter SuppressionClutter Suppression Performance EvaluationCompoundingSignalsUltrasonic Imaging
One of the key limitations in ultrasonic imaging results from clutter signals corresponding to a large number of coherent, unresolved and randomly distributed reflectors located around the desired target. The backscattered echoes from such reflectors can severely deteriorate the image quality to the point of concealing the desired target. This fundamental problem which exists in many imaging and detection problems cannot be eliminated by conventional noise reduction techniques such as time averaging. However, in recent years clutter suppression has been achieved in A-scan applications using some novel frequency and spatial compounding techniques, where the clutter signals are decorrelated by shifting the transmitted frequency or the transducer position, respectively, prior to processing. In the work presented, these techniques are extended to imaging applications and evaluated using experimental data obtained from large grained samples. The experimental results indicate that the frequency and spatial compounding techniques are capable of improving the image quality beyond the conventional averaging techniques.
10
Table of Integrals, Series and Products.
Y. L., I. S. Gradshteyn, I. M. Ryzhik · Mathematics of Computation · 1981 · 8.6K citations
Ralph G. Hudson · Medical Entomology and Zoology · 1917 · 7K citations
Natural Sciences, Computer Science, Differential Equations +7
Acoustic speckle: Theory and experimental analysis
John G. Abbott · Ultrasonic Imaging · 1979 · 269 citations