IEEE Transactions on Medical Imaging · 2003 · 50 citations · 14 references
MedicineBioengineeringBiomedical EngineeringDiagnostic ImagingUltrasonographyUltrasoundImage RegistrationMedical ImagingThree-dimensional ReconstructionThree-dimensional UltrasonographyThree-dimensional UltrasoundUltrasound ImagingFreehand Three-dimensional UltrasoundNonrigid Image RegistrationPrecisionRegistrationRegistration Of Scan DataSpline ModelingThree-dimensional Ultrasound Imaging
A freehand scanning protocol is the only way to acquire arbitrary large volumes of three-dimensional ultrasound (US) data. For some applications, multiple freehand sweeps are required to cover the area of interest. Aligning these multiple sweeps is difficult, typically requiring nonrigid image-based registration as well as the readings from the spatial locator attached to the US probe. Conventionally, nonrigid warps are achieved through general elastic spline deformations, which are expensive to compute and difficult to constrain. This paper presents an alternative registration technique, where the warp's degrees of freedom are carefully linked to the mechanics of the freehand scanning process. The technique is assessed through an extensive series of in vivo experiments, which reveal a registration precision of a few pixels with comparatively little computational load.
14
Rapid calibration for 3-D freehand ultrasound
Richard W. Prager, Robert Rohling, Andrew H. Gee et al. · Ultrasound in Medicine & Biology · 1998 · 386 citations
High-definition freehand 3-D ultrasound
Graham M. Treece, Andrew H. Gee, Richard W. Prager et al. · Ultrasound in Medicine & Biology · 2003 · 157 citations