Analysis of Early Failure of Tunneled Hemodialysis Catheters

37

Citations

20

References

2002

Year

Abstract

By paying careful attention to catheter tip position, searching diligently for the presence of a fibrin sheath when catheter exchanges are made over a wire, and better investigating presumed catheter infection, we could reduce the early failure rate by more than half, from 46 cases to 20 cases (nine cases of suspected infection were in fact not infected).