DYNAMIC BEAM BASED CALIBRATION OF BEAM POSITION MONITORS

Bernd Dehning, F. Tecker, G. Ferri, J. Wenninger, Peter S. Galbraith, M. Placidi, F. Sonnemann, G. Mugnai

International Linear Collider · 1998 · 10 citations · 1 references

Full text

Open access

Concepts

Abstract

The degree of spin polarization at LEP is strongly dependent on the knowledge of the vertical orbit. Quadrupole magnet alignment and beam position monitor (BPM) offsets are the main source of the orbi t uncertainty. The error of the orbit monitor readings can be largely reduced by calibrating the monitor relative to the adjacent quadrupole. At LEP, 16 BPM offsets can be determined in parallel durin g 40 minutes. The error of the measure offset is about 30mm. During the LEP run 1997, more than 500 measurements were made and used for the optimisation of polarization. The method of dynamic beam bas ed calibration will be explained and the results will be shown.

References

1