International Linear Collider · 1998 · 10 citations · 1 references
Open access
EngineeringNuclear EngineeringInstrumentationNuclear InstrumentationRadiation MeasurementCalibrationOrbit DeterminationBeam Position MonitoringDynamic Beam AnalysisPolarimetric CalibrationPolarization MeasurementsSpin Polarization
The degree of spin polarization at LEP is strongly dependent on the knowledge of the vertical orbit. Quadrupole magnet alignment and beam position monitor (BPM) offsets are the main source of the orbi t uncertainty. The error of the orbit monitor readings can be largely reduced by calibrating the monitor relative to the adjacent quadrupole. At LEP, 16 BPM offsets can be determined in parallel durin g 40 minutes. The error of the measure offset is about 30mm. During the LEP run 1997, more than 500 measurements were made and used for the optimisation of polarization. The method of dynamic beam bas ed calibration will be explained and the results will be shown.
1
I. Barnett, A. Beuret, B. Dehning et al. · AIP conference proceedings · 1995 · 12 citations · Full text