IEEE Transactions on Biomedical Engineering · 1995 · 54 citations · 15 references
EngineeringCardiologyPhysiologyBiomedical Signal ProcessingElectrophysiologyExercise PhysiologyArtifact RejectionBiomedical Signal FilteringCardiac Signal ProcessingCardiovascular MonitoringDigital Signal ProcessingElectrocardiographyEcg Signal DenoisingElectrocardiographic Artifact AnalysisFilter Performance EvaluationFourier Series AnalysisHeart RateNoise Modelling
Impedance cardiography (ICG) may be altered by noises as respiration and movement artifacts, mainly during exercise. In this work, a scaled Fourier linear combiner (SFLC) event-related to the R-R interval of ECG is proposed. It estimates the deterministic component of the impedance cardiographic signal and removes the noises uncorrelated to this interval. The impedance cardiographic signal is modeled as Fourier series with the coefficients estimated by the least mean square (LMS) algorithm. Simulations have been carried out to evaluate the filter performance for different noise conditions. Moreover, the method capability to remove uncorrelated noises was also examined in physiological data obtained in rest and exercise, by synchronizing respiration and pedaling with a metronome. Analyzing the ICG power spectrum, it was concluded that the proposed filter could remove the noises that are not synchronized with heart rate.
15
Theory and application of digital signal processing
T.N. Trick · Proceedings of the IEEE · 1975 · 1.9K citations
Development and evaluation of an impedance cardiac output system.
W. G. Kubicek, James N. Karnegis, Robert Patterson et al. · PubMed · 1966 · 1.1K citations