Concepedia

Publication | Closed Access

Comparison of Field Performance of the Andersen N6 Single Stage and the SAS Sampler for Airborne Fungal Propagules

25

Citations

0

References

2001

Year

Abstract

Comparisons of two common bioaerosol samplers were made after sampling and enumeration of airborne fungal propagules in several office structures on a university campus in Southern California. Data collected on five occasions throughout the year showed that a Surface Air Systems (SAS) high flow portable sampler recovered consistently lower levels of colony forming units (cfu) than an Andersen N6 single stage impactor. There was no difference statistically between the samplers when concentrations of Cladosporium were compared. Compared to the Andersen N6, the SAS sampler recovered about half the number of cfu for three other fungal categories, i.e. non-sporulating species, Aspergillus and Penicillium and others. Differences in sampler efficiencies are discussed in terms of effective particle diameters. Counts of culturable airborne fungal spores obtained with the SAS sampler should be interpreted with caution when genera other than Cladosporium predominate.