Pacing and Clinical Electrophysiology · 2011 · 14 citations · 4 references
MedicineCardiologyCardiothoracic SurgeryClinical CardiologyElectrophysiologyCardiac ArrhythmiasCardiac AssessmentCardiac Risk AssessmentConduction DisordersElectrocardiographyVentricular ElectrophysiologyCardiac Conduction System PathologyConduction Abnormality AnalysisConduction Disorder DetectionElectrocardiographic Lead SystemsInsulation Failure DetectionProvocative Clinical Tests
We report a patient with ischemic cardiomyopathy status post implantable cardioverter-defibrillator (ICD) implantation in June 2009. Outer insulation failure of her Riata ST ventricular ICD lead goes undetected by routine interrogation and provocative tests as recommended from its manufacturer.(1) Another domain of assessment, fluoroscopy, to facilitate early detection of structural abnormality in this family of ICD lead is discussed.