Dual-energy contrast-enhanced digital mammography: patient radiation dose estimation using a Monte Carlo code

E. Yakoumakis, E. Tzamicha, A. Dimitriadis, Evangelos Georgiou, Virginia Tsapaki, Athanasios N. Chalazonitis

Radiation Protection Dosimetry · 2015 · 11 citations · 10 references

Concepts

Abstract

Mammography is a standard procedure that facilitates breast cancer detection. Initial results of contrast-enhanced digital mammography (CEDM) are promising. The purpose of this study is to assess the CEDM radiation dose using a Monte Carlo code. EGSnrc MC code was used to simulate the interaction of photons with matter and estimate the glandular dose (Dg). A voxel female human phantom with a 2-8-cm breast thickness range and a breast glandular composition of 50 % was applied. Dg values ranged between 0.96 and 1.45 mGy (low and high energy). Dg values for a breast thickness of 5.0 cm and a glandular fraction of 50 % for craniocaudal and mediolateral oblique view were 1.12 (low energy image contribution is 0.98 mGy) and 1.07 (low energy image contribution is 0.95 mGy), respectively. The low kV part of CEDM is the main contributor to total glandular breast dose.