IEEE Transactions on Instrumentation and Measurement · 1993 · 23 citations · 4 references
EngineeringElectrical EngineeringHeat TransferPower ElectronicsThermal AnalysisAc-dc ConversionAc-dc ConvertersHeat ConductionHeat Transfer AnalysisTemperature MeasurementThermal ManagementComparator DesignHeat Flux MeasurementMeasurement System DesignMeasurement System ImplementationMeasurement System VerificationTemperature ConversionThermal TransferTransfer Standards
A new measurement system is developed in order to re-evaluate the present AC-DC transfer standards of the Electrotechnical Laboratory (ETL). The measurement system consists of a high-precision comparator system and a modified fast-reversed DC source. The AC-DC transfer differences of thermal transfer standards are compared using the comparator system with a resolution better than 1 ppm. The absolute AC-DC transfer differences of single-junction thermal converters (SJTCs) are evaluated in term of a multijunction thermal converter and by the modified fast-reversed DC method. The results obtained by the two methods agree within 1 ppm.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
4
F. L. Hermach, Joseph R. Kinard, John R. Hastings · IEEE Transactions on Instrumentation and Measurement · 1987 · 37 citations · Full text