In this study, an analysis of biological calibration was performed and the presented results included calibration error, calibration time, and calibration difficulty. These values determine potential applications of Web-based hearing tests conducted in home settings and are decisive factors when selecting the calibration method. If there are no substantial time limitations, it is advisable to use Bekesy method and determine the reference sound level independently at each frequency because this approach is characterized by the lowest error.