IEEE Transactions on Nuclear Science · 1995 · 16 citations · 23 references
EngineeringPhysicsRadiologyDiagnostic ImagingRadiation DetectionRadiation MeasurementDetector TechnologyRadiation DetectorsRadiation Imaging DetectorsRadiographic ImagingX-ray DetectionX-ray DetectorsHigh-resolution X-ray DetectorsLinearityMultiwire Proportional ChambersPosition-sensitive DetectorsTwo-dimensional ImagingX-ray Imaging PerformanceX-ray Imaging PhysicsX-ray Imaging Systems
A method of construction of multiwire chambers has been developed which yields X-ray images with very small non-linearity, in particular that due to anode wire modulation. The technique requires an absorption region in the detector, with the anode and upper cathode wires, which are parallel to one another, in registration with each other. Studies show that the strength of the electric field in the drift region determines the degree of anode wire modulation and, under specific conditions, modulation is halved in period and significantly reduced in amplitude. New interpolating cathodes have been developed to read out position in the other axis, along the anode wire direction. Position resolution in both axes-between 100-150 /spl mu/m FWHM has been achieved. Two-dimensional images of small objects have been taken which show unprecedented linearity.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
23
G.P. Yost, R. M. Barnett, I. Hinchliffe et al. · Physics Letters B · 1988 · 4.4K citations
Hidehiro Iida, Iwao Kanno, Shuichi Miura et al. · Journal of Cerebral Blood Flow & Metabolism · 1986 · 376 citations · Full text
Continuous-slice PENN-PET: a positron tomograph with volume imaging capability.
Joel S. Karp, G. Muehllehner, David A. Mankoff et al. · PubMed · 1990 · 237 citations