Patient Alert, based on daily lead-impedance measurement, detected one-third of all ICD lead failures. Combined use with continuous lead integrity monitoring (SIC) increased sensitivity to 100%. Integrated bipolar leads may yield a false-positive SIC. Incorporating SIC and automated pace/sense threshold measurement may improve Patient Alert sensitivity for severe lead complications.