Radiation Protection Dosimetry · 2008 · 24 citations · 0 references
Natural SciencesNuclear EngineeringNuclear MedicineNuclear PhysicsRadiation DetectionRadiation DosimetryAbsorbed DoseAlpha DecayNuclear DecayRadiation DetectorsRadiation Dose MeasurementRadioactive DecayDetectorsIrradiation DamageRadium Exposure AssessmentRadonRadon GasRadon MeasurementRadon Progeny DetectionRadon Risk Assessment
The new RAMARN system for radon volume activity measurement was developed in 2003 and has been in use since then. RAMARN system consists of a plastic chamber that is conically cylindrical in shape and about 0.5 l in volume; a bare Kodak LR 115 is located on the bottom of this diffusion chamber. The size was chosen to avoid the influence of deposited decay products of radon. Kodak has a spectrometric character--the tracks are visualized only for alphas with energies between 1 and 3 MeV that touch the foil; thus the effective volume has a lens shape. The response therefore corresponds to diffused radon and half of (218)Po born by radon gas decay. The experiments described below were conducted as one part of a routine methodology control, focused on classifying worker irradiation from natural ionizing radiation sources in show caves and in caves used for speleotherapy.