Bioanalysis · 2015 · 73 citations · 16 references
PharmacyBiomedical EngineeringPharmaceutical AnalysisAnalytical ChemistryElectroanalytical MethodsInstrumentationBlood StorageSample CollectionSample HandlingSample PreparationBioanalytical Sample HandlingBlood Sample CollectionFingerstick SamplingLiquid Volume MeasurementSample Collection SystemsSample Volume MeasurementSamplingWhole Blood Collection
BACKGROUND: DBS samples collected from a fingerprick typically vary in volume and homogeneity and hence make an accurate quantitative analysis of DBS samples difficult. RESULTS: We report a prototype which first defines a precise liquid volume and subsequently stores it to a conventional DBS matrix. Liquid volumes of 2.2 µl ± 7.1% (n = 21) for deionized water and 6.1 µl ± 8.8% (n = 15) for whole blood have been successfully metered and stored in DBS paper. CONCLUSION: The new method of collecting a defined volume of blood by DBS sampling has the potential to reduce assay bias for the quantitative evaluation of DBS samples while maintaining the simplicity of conventional DBS sampling.
16