Journal of Clinical Pathology · 2012 · 16 citations · 11 references
Accuracy of CoaguChek XS INR determinations can be assessed by a group of collaborating centres using a limited number of capillary blood samples. As the mean INR differences with the International Standard were smaller than 10%, the lots were approved for use by the Netherlands Thrombosis Services.
11
Herbert Watzke, Michael Hasenkam, Michael Schwab et al. · Thrombosis and Haemostasis · 2007 · 49 citations