MedicineBiomedical EngineeringRadiologyDiagnostic ImagingDiagnostic RadiologyNeuroradiologyBrain ImagingCerebrovascular ImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsNeuroimaging PhysicsNeurointerventional ImagingIn-vitro CharacterizationIntracranial Aneurysm ImagingMagnetic Field StrengthMagnetic Resonance Imaging AcquisitionRadiofrequency Coil DesignRadiofrequency CoilsSafety In Magnetic Resonance ImagingSusceptibility Effects
A detachable coil was assessed for magnetic field interactions, heating (body RF coil, SAR of 2.0-W/kg), and artifacts at 3-Tesla. The coil showed no magnetic field interactions and heating was negligible (+0.3 degrees C). Therefore, this coil is safe for a patient undergoing MR imaging at 3T or less. While artifacts may impair the ability to properly visualize anatomy in proximity to this implant, careful selection of imaging parameters can mitigate this problem.