MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyDetector TechnologyGamma-ray DetectorsGamma-ray InstrumentationImage QualityImage Quality AssessmentRadiographic ImagingCamera SystemsGamma Camera ImagingGamma Camera SystemsGamma DosimetryScintillator CrystalsSingle-photon Emission Computed Tomography Detectors
The performances of two commerical gamma cameras, with crystals 1.3 and 0.6 cm thick, are compared. Phantom studies simulating clinical conditions showed no significant difference in performance at 140 keV. At 68 keV, the thinner crystal gave marginal improvements in camera performance with phantoms simulating clinical conditions. Frequent use with very low-energy emitters, such as Tl-201, would be needed to justify the expense required to refit the 1.3 cm camera with a 0.6 cm crystal.