A study of the constant-temperature hot-wire anemometer
99
Citations
2
References
1971
Year
EngineeringStray Bridge ReactanceMeasurementCalibrationBridge ImbalanceTemperature MeasurementConstant-temperature Hot-wire AnemometerEducationDynamic ResponseHeat TransferInstrumentationThermal SensorThermal Engineering
The conventional ‘bridge-feedback amplifier’ constant-temperature hot-wire anemometer is analysed to determine its static and dynamic response. The effects of moderate feedback amplifier gain, bridge imbalance, stray bridge reactance, amplifier offset voltage, lack of common mode rejection, amplifier frequency response and departure from constant transconducture are included. The root loci of the system are mapped out and the consequences of the analysis are discussed from the viewpoint of both the operator and the designer.