Publication | Open Access
Laboratory Evaluation of Four Different Devices for Secretion Mobilization: Acapella Choice, Green and Blue Versus Water Bottle
16
Citations
8
References
2013
Year
Setting 4 was the best for all 3 Acapella devices, and the filling height of the water bottle should be 5 cm. At these settings, all devices elicited vibration frequencies between 12 and 15 Hz, which is theoretically optimal for secretion mobilization. The resistance pressures of the devices to elicit these vibrations were between 5 and 11 cm H2O. However, the Acapella devices elicit higher vibration amplitudes (5-8 cm H2O) than the water bottle (1.8 cm H2O) CONCLUSIONS:: Setting 4 was optimal for all 3 Acapella devices. The Acapella devices may be more efficient for secretion mobilization than the water bottle, because they elicit greater amplitude of vibrations.