A 5-laboratory study was performed that used the National Committee for Clinical Laboratory Standards (NCCLS) reference agar dilution method with 3 media formulations to determine whether the use of different media would affect minimum inhibitory concentration (MIC) results. Wilkins-Chalgren, Brucella-based blood agar (BRU), and Wilkins-Chalgren agar plus blood (WCB) and 6 antibiotics (clindamycin, cefoxitin, ceftizoxime, piperacillin, metronidazole, and trovafloxacin) were evaluated with 58 isolates. The MIC values were compared, and a significant correlation of >0.80 was demonstrated for all media and each antibiotic/organism group. The cumulative rate of errors for all antibiotics was 0.1%. These data indicate that a change in the NCCLS reference medium for testing of anaerobic bacteria susceptibility to either BRU or WCB will not affect the MIC results for the antibiotics and organisms evaluated.