Journal of neurosurgery · 1985 · 15 citations · 9 references
MedicineCardiologyClinical CardiologyMedical PhysicsSurgical InstrumentationBiopotential MeasurementCardiovascular MonitoringExperimental Model TestingShunt MalfunctionShunt PhysiologySurgical ShuntingThermistorsVentricular Catheter MonitoringVentriculoperitoneal Shunt Surgery
A simple method is described for checking the patency of ventriculoperitoneal shunts by using a pair of small disc thermistors. With this technique, shunt patency is shown by a downward deflection of the recording trace, blockage of the ventricular catheter is indicated by a flat-line recording, and obstruction of the distal end of the abdominal catheter is demonstrated by an upward deflection. These observations were confirmed in a study of 32 hydrocephalic patients with obstructed shunts who subsequently underwent shuntography and operation. An experimental model was also devised to study the mechanism of the recording changes.
9
Radionuclide assessment of cerebrospinal fluid shunt function in children.
Thomas G. Rudd, David Shurtleff, John D. Loeser et al. · PubMed · 1973 · 50 citations
Medicine, Pediatrics, Radiology +11
A simple test of shunt function: the shuntgram
Richard C. Dewey, Edward J. Kosnik, Martin P. Sayers · Journal of neurosurgery · 1976 · 35 citations
Medicine, Cardiology, Radiology +8