Matching of adjacent radiation beams for isocentric radiotherapy

M. P. Casebow

British Journal of Radiology · 1984 · 24 citations · 6 references

Concepts

Abstract

In radiotherapy, it is sometimes necessary to treat large or awkwardly shaped structures with two or more adjacent radiation beams, rather than a single beam. An even dosage across the junction between two fields may be obtained for all depths if the fields are orientated so that the planes representing their adjacent edges (e.g., the 50% decrement) are made to coincide. Half-beam lead blocks have been used in this way to match orthogonal fields, where the beam edges are butted together in a vertical “match plane” (Williamson, 1979). Unfortunately half-beam blocks reduce the length of the fields available, and moreover, as they cover a large area they must either be heavy or allow a substantial penetration of the beam. Some radiotherapy machines allow one collimator block to be moved to the centre to produce this form of treatment. Apart from beam shaping, the half-beam block is used to orientate the beam in a manner additional to that provided by the gantry angle. This effect may, however, be obtained by suitable angulation of the gantry, collimators and treatment table (Fleming & Orchard, 1974; Casebow, 1976; Siddon, 1981). The equations describing these set-up angles are derived below for the general case of matching adjacent fields without the use of lead blocks, and their application is then illustrated using the example of treatment for breast carcinoma and medulloblastoma.