Journal of Periodontal Research · 1988 · 86 citations · 12 references
DentistryOral BiologyPeriodontologyDental MaterialsPeriodontal BiologyPeriodontal MedicineChromatographyFluid AnalysisGingival BiologyAbsorbent MaterialsCalibration Curve ApproachCalibration Curve ConstructionCalibration Curve FittingFilter DesignGingival Crevicular Fluid SamplingPaper ChromatographyPaper-based Analytical Devices
Four different paper types were evaluated in vitro for their suitability for the collection of gingival crevicular fluid and its subsequent quantification using both the Periotron® 6000 and measurement of the area of paper wetted. Whatman 3MM chromatography paper (Wh.3MM) was found to be the most suitable and strip dimensions of 2x8 mm were found to be the most absorbent whilst remaining compatible with the dimensions of the Periotron electrodes. The 2×8 mm Wh.3MM paper strips were used to derive a quadratic calibration curve, for a range of fluid volumes, which was shown to be a better fit than a straight line. Recovery of human serum proteins applied to these strips exceeded 90% even after compression of the strip between the electrodes of the Periotron during fluid volume measurement.
12
Gingival crevicular fluid antibody to oral microorganisms
Jeffrey L. Ebersole, Martin A. Taubman, Daniel J. Smith et al. · Journal of Periodontal Research · 1984 · 103 citations
The Evaluation of the Periotron
Jerry J. Garnick, Robert Pearson, Dan Harrell · Journal of Periodontology · 1979 · 47 citations