Concepedia

Publication | Closed Access

Preservation of residual hearing with cochlear implantation: How and why

275

Citations

15

References

2004

Year

Abstract

Hearing threshold level data were available for 12 patients recruited from 6 of the centres. Median increases in hearing threshold levels were 23, 27 and 33 dB for the frequencies 125, 250 and 500 Hz, respectively. These median increases include the data for two patients who had total loss of residual hearing due to difficulties encountered during surgery. "Cochlear view" X-ray images indicated that the depth of insertion varied between 300 and 430 degrees, despite modest variations in the length of the electrode inserted (17-19 mm). The insertion angle had some influence on the preservation of residual hearing at frequencies of 250-500 Hz. Six of the 12 patients retained sufficient hearing for effective use of an ipsilateral ITE hearing aid (< or = 80 dB HL at 125 and 250 Hz; < or = 90 dB HL at 500 Hz). Word recognition scores in quiet were improved from 10% to 30% with the cochlear implant plus ipsilateral hearing aid in 3 patients who had at least 3 months postoperative experience. Signal:noise ratio thresholds for sentence recognition were improved by up to 3 dB. Patients reported that they experienced greatly improved sound quality and preferred to use the two devices together.

References

YearCitations