IEEE Transactions on Instrumentation and Measurement · 2008 · 25 citations · 14 references
EngineeringElectrical EngineeringInstrumentationOptoelectronicsPhotonicsCalibrationMicrowave MeasurementsOptical InstrumentationUltrafast OpticsImpulse ResponseMonte Carlo AnalysisOscilloscope CalibrationOscilloscope-based MeasurementRise TimeRise Time MeasurementStep ResponseTime-domain Response Analysis
We report on the enhancement of the Physikalisch-Technische Bundesanstalt's (PTB's) ultrafast optoelectronic measurement facility by characterizing the full impulse and step response of a 70-GHz sampling oscilloscope. A novel optoelectronic technique for the generation and the detection of ultrashort voltage pulses, which serve as calibration signals for the oscilloscope, is introduced. The uncertainty of the oscilloscope's time-domain response is derived from a Monte Carlo analysis. Aside from a more complete characterization, our enhanced technique considerably reduces the uncertainty of rise-time measurements.
14
7 terahertz broadband GaP electro-optic sensor
Qing Yang Steve Wu, Xicheng Zhang · Applied Physics Letters · 1997 · 351 citations
LT-GaAs detector with 451 fs response at 1.55 µmvia two-photon absorption
H. Erlig, S. Wang, Talal Azfar et al. · Electronics Letters · 1999 · 52 citations