Journal of Diabetes Science and Technology · 2010 · 39 citations · 28 references
MedicineCritical Care NursingDiabetologyHealth AssessmentPatient SafetyBlood Glucose MonitoringContinuous Glucose MonitoringDiabetes MellitusGlycemic ControlGlycemic MonitoringMetabolic MonitoringAccuracyBlood Glucose Meter AccuracyBlood Glucose RegulationDiabetic Ketoacidosis AssessmentGlucose Analysis In Body FluidsGlucose MonitoringGlucose Monitoring DevicesInpatient Glycemic Control
BACKGROUND: Glucose management in an intensive care unit (ICU) is labor-intensive. A continuous glucose monitoring system (CGMS) has the potential to improve efficiency and safety in this setting. The goal of this study was to determine if the Medtronic Guardian REAL-Time CGMS was accurate and tolerated by patients in a rural hospital ICU unit. METHOD: Differences between individual finger stick blood glucose (FSBG) and CGMS values were compared to American Diabetes Association (ADA) and International Organization for Standardization (ISO) standards. Continuous glucose monitoring system accuracy was evaluated over four ranges: <75, 75-140, 140-200, and >200 mg/dl. Other accuracy measures [mean absolute deviation (MAD), mean absolute relative difference (MARD), and coefficient of linear regression of CGMS on FSBG] were calculated. Nursing staff and patients were surveyed regarding use of the CGMS in the ICU. RESULTS: Twenty-nine participants had 320 FSBG and corresponding CGMS readings. Sixty-two percent of participants were admitted with diabetic ketoacidosis (DKA). Two hundred and thirteen (66.6%) were accurate within the ISO standard, whereas only 70 out of 320 (21.9%) were within the 5% ADA standard. The CGMS was most accurate in euglycemia. Technical difficulties, such as adequate time for "wetting" and calibration of electrodes, arose with the sensors. The MAD was 28.3 mg/dl, the MRAD was 17.4%, and the linear regression coefficient of CGMS on FSBG was 0.834 (p < 0.001). CONCLUSIONS: The CGMS is well tolerated by ICU patients but, at present, is not sufficiently accurate to be used for therapeutic decisions in the acute setting, particularly in patients with diabetic ketoacidosis. There is a need to find resolution to the technical issues regarding electrode "wetting" and calibration if CGMS use in the ICU setting is to provide an effective means of diabetes care and management.
28
Intensive Insulin Therapy in Critically Ill Patients
Greet Van den Berghe, Pieter Wouters, Frank Weekers et al. · New England Journal of Medicine · 2001 · 10K citations · Full text
Standards of Medical Care in Diabetes—2010
Diabetes Care · 2009 · 3.5K citations · Full text
Evaluating Clinical Accuracy of Systems for Self-Monitoring of Blood Glucose
William L. Clarke, Daniel J. Cox, Linda Gonder‐Frederick et al. · Diabetes Care · 1987 · 1.4K citations
Medicine, Endocrinology, Biosensors +16