IEEE Transactions on Ultrasonics Ferroelectrics and Frequency Control · 2007 · 51 citations · 44 references
Veterinary MedicineBioengineeringBiomedical EngineeringBiomedical ImagingDiagnostic ImagingUltrasonographyDiagnostic UltrasonographyElastographyStrain AnalysisUltrasonicsUltrasound ImagingUltrasound PhysicsAmplitude VariationsDisplacementMeasurement Noise EstimationNoise EstimationNoise-corrected EstimationUltrasonic ElastographyUltrasonic Imaging
Ultrasonic strain imaging usually begins with displacement estimates computed using finite-length sections of RF ultrasound signals. Amplitude variations in the ultrasound are known to perturb the location at which the displacement estimate is valid. If this goes uncorrected, it is a significant source of estimation noise, which is amplified when displacement fields are converted into strain images. We present a study of this effect based on theoretical analysis and practical experiments. A correction method based on the analysis is tested on phase zero and correlation coefficient strain imaging, and compared to the amplitude compression techniques of earlier studies. We also test adaptive strain estimation to provide a benchmark, but the performance of our new method matches or surpasses this benchmark under normal scanning conditions. Furthermore, the new correction is suitable for real time applications owing to its extreme computational simplicity.
44
Elastography: A Quantitative Method for Imaging the Elasticity of Biological Tissues
J. Ophir, I. Céspedes, H. Ponnekanti et al. · Ultrasonic Imaging · 1991 · 3.5K citations
Coherence and time delay estimation
G.C. Carter · Proceedings of the IEEE · 1987 · 1.2K citations