Highly sensitive and specific HPLC with fluorometric detection for determination of plasma epinephrine and norepinephrine applied to kinetic studies in humans

Jacques J. Willemsen, H A Ross, M.‐C. Jacobs, Jacques W.M. Lenders, Th. Thien, L M Swinkels, T J Benraad

Clinical Chemistry · 1995 · 93 citations · 12 references

Concepts

Abstract

An HPLC separation method combined with fluorometric detection was extended to enable simultaneous assessment of plasma 3H-labeled and endogenous epinephrine (E) and norepinephrine (NE). Forearm fractional extraction (FFE) of 3H-labeled E and NE and of endogenous E was measured in 40 healthy volunteers who were receiving a continuous infusion of 3H-labeled E and NE. Concentrations of arterial and venous E were 26.8 +/- 1.95 (mean +/- SE) and 6.8 +/- 0.75 ng/L, respectively. Arterial and venous NE and dopamine (DA) were also measured, with respective values of 140.7 +/- 8.5 and 192.1 +/- 15.1 for NE, and 13.1 +/- 0.78 and 11.3 +/- 0.70 ng/L for DA. The FFE of 3H-labeled E was slightly but significantly higher (0.790 +/- 0.016) than the that of either 3H-labeled NE or endogenous E (0.748 +/- 0.0146 and 0.745 +/- 0.0185, respectively; P < 0.001), the correlations being highly significant (r = 0.80, P < 0.001) in both cases. The small difference between the FFE of E and of 3H-labeled E allows the calculation of the apparent spillover of E. However, this spillover was negligible compared with forearm NE spillover (0.0112 +/- 0.0031 vs 1.369 +/- 0.128 ng/L per minute. The high sensitivity of this measurement of venous E widens the possibilities for studying E kinetics under physiological conditions.

References

12