Laboratory Animals · 1988 · 78 citations · 6 references
Natural SciencesAnimal SciencesAcousticsUltrasonographyUltrasoundEnvironmental AcousticsEnvironmental Noise AssessmentNoise ExposureNoise MeasurementUltrasonicsUltrasound PhysicsAcoustic EnvironmentsNoise Source CharacterizationSound Pressure LevelUltrasonic AcousticsUltrasonic WavesUltrasound Safety
Many laboratory animals are known to be sensitive to sounds (ultrasounds) beyond the nominal upper limit (20 kHz) of the human hearing range. Sources of sound in laboratories and animal houses were examined to determine the extent of ambient ultrasound. Of 39 sources monitored, 24 were found to emit ultrasonic sounds. Many of these (e.g. cage washers and hoses) also produced sound in the audible range. Running taps, squeaky chairs and rotating glass stoppers created particularly high sound pressure levels and contained frequencies to over 100 kHz. The oscilloscopes and visual display units investigated provided particular cause for concern as they emitted sounds that were entirely ultrasonic and therefore were apparently silent. Ambient ultrasound therefore appears to be common in laboratories and animal houses. It is suggested that its effect on laboratory animals should be investigated and guidelines on acceptable levels be formulated.
6
Development of absolute auditory thresholds in the house mouse (Mus musculus).
Günter Ehret · PubMed · 1976 · 215 citations
Life Sciences, Neuroscience, Zoology +17
Hearing range of the domestic cat
Rickye S. Heffner, Henry E. Heffner · Hearing Research · 1985 · 178 citations
Hearing in the Elephant ( <i>Elephas maximus</i> )
Rickye S. Heffner, Henry E. Heffner · Science · 1980 · 82 citations