IEEE Transactions on Nuclear Science · 1990 · 58 citations · 1 references
MedicineNuclear MedicineRadiologyDiagnostic ImagingDiagnostic Nuclear MedicineDiagnostic RadiologyDetector TechnologyDiagnostic Imaging SystemsGamma-ray DetectorsGamma-ray InstrumentationMedical ImagingRadiation Imaging DetectorsCircuit ConfigurationDetector DesignDetector Spatial ResolutionGamma-ray ImagingGeometric Distortion CorrectionLinearityPhotomultiplier TubesPosition-sensitive Detectors
It is pointed out that many potentially practical applications for nuclear medicine diagnostic procedures are difficult or impossible to perform because of the physical size of existing mobile gamma cameras. With the commercial availability of position sensitive photomultiplier tubes (PSPMTs) the development of an instrument ideally suited to these applications now appears to be practical. The major components necessary to construct such a camera have been tested, and a circuit configuration designed to provide a high level of clinical performance in such a device is proposed. Spatial resolution and linearity data from a developmental prototype are provided. Measurements have shown that by using conventional sum and difference analog circuits only, a 76-mm/sup 2/ PSPMT has good linearity and therefore uniformity of sensitivity over only the central 60% of its absolute field of view. It is believed that this high-quality imaging area can be increased to virtually the full field of view by implementing a digital geometric correction scheme using a position vector lookup table.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
1
Hal O. Anger · Review of Scientific Instruments · 1958 · 791 citations