Self‐shading of in‐water optical instruments

Howard R. Gordon, Kuiyuan Ding

Limnology and Oceanography · 1992 · 222 citations · 17 references

Concepts

TL;DR

The study proposes a simple correction model for self‑shading error that relaxes size constraints on in‑water optical radiometers. Self‑shading was quantified by Monte Carlo simulations of the light field with and without the radiometer. The simulations show that self‑shading error depends on instrument size and water absorption, is largely independent of scattering, and can exceed 15–20 % unless the radiometer diameter is limited to ≲ 1/30 a for irradiance and ≲ 1/100 a for upwelling; the proposed correction reduces these limits to ≲ 1/6 a and ≲ 1/30 a, while error remains high at high pigment concentrations or wavelengths above 600 nm.

Abstract

The self‐shading of in‐water optical radiometers for measuring upwelling ( L u ) and irradiance ( E u ) is estimated from Monte Carlo simulations of the light field in the presence and absence of the instrument. It is found that the error ( ε ) induced by the presence of the radiometer is a function of its size and the absorption coefficient ( a ) of the medium, i.e. when ε 15–20% the error is independent of the scattering coefficient ( b ). Direct measurement of L u or E u with ε ≤ 5% places severe limitations on the instrument size, e.g. in the case of a cylindrical housing and small solar zenith angles (the worst case) the diameter of the instrument must be ≲ 1/30 a for E u and ≲ 1/100 a for L u . A correction method based on a simple model of ε is proposed and it is shown that when used these constraints on the diameter are reduced to 1/6 a and 1/30 a , respectively, for an after‐correction error of 5%. The self‐shading error is estimated across the spectrum as a function of the pigment concentration of case 1 waters, and it is found that it can be large for typical radiometers, especially at high pigment concentrations or wavelengths >600 nm.

References

17