Publication | Closed Access
Total nucleated cell differential for blood and bone marrow using a single tube in a five‐color flow cytometer
52
Citations
33
References
2007
Year
We have designed a one-tube immunophenotyping panel for classification of total nucleated cells and platelets in blood or bone marrow. The seven parameters available in one single tube in our cytometer seem to be enough for reliable differential count even in difficult pathological samples. The analytical imprecision of the flow cytometer differential was much lower than that obtained with microscopy or cell counter differentials.