Design changes in continuous-flow left ventricular assist devices and life-threatening pump malfunctions

33

Citations

24

References

2014

Year

Abstract

The modified cable strain relief of the HM II pump and the sintering of the inflow cannula of the HW pump demonstrated a significant reduction in the incidence of life-threatening pump-related complications, whereas the sealed inflow connector and outflow graft seem to be associated with a higher incidence of pump thrombosis. However, the overall incidence of pump-related complications after the latest design changes was similar for both pumps over a 2.5-year period.