Concepedia

Publication | Open Access

Accuracy of Doppler-echocardiographic parameters for the detection of aortic bileaflet mechanical prosthetic valve dysfunction

17

Citations

18

References

2013

Year

Abstract

The Doppler-echocardiographic parameters and criteria proposed in the guidelines lack sensitivity for the detection of BMV dysfunction. The utilization of a DVI < 0.35 or an EOA-D > 1 SD improved the sensitivity (>90%) for the detection of moderate-to-severe dysfunction, but the sensitivity remained suboptimal (<65%) for detection of mild dysfunction.

References

YearCitations