IEEE Transactions on Applied Superconductivity · 1995 · 10 citations · 4 references
Natural SciencesElectrical EngineeringPhysicsCondensed Matter PhysicsElectromagnetismMagnetic MaterialsMagnetic CharacterizationMagnetic Thin FilmsSuperconducting Thin FilmsThin Film Magnetic PropertiesJosephson JunctionsJosephson TunnelingMagnetic GradiometersMagnetic GradiometryMagnetometer InstrumentationMagnetometersSquid MagnetometersWhite Noise
Thin film DC SQUID gradiometers with a large effective pickup area using only a single YBa/sub 2/Cu/sub 3/O/sub 7-x/ layer and step edge Josephson junctions on SrTiO/sub 3/ substrate have been prepared. The planar gradiometer consist of two loops coupled directly to a small area SQUID in the center of the structure. The main advantage in comparison to magnetometers is the possibility of direct measurements in magnetically unshielded environment. At 77 K without bias reversal technique we obtain a white noise level about 3.6/spl times/10/sup -5/ /spl Phi//sub 0//Hz/sup 1/2/ or 700 fT/cmHz/sup 1/2/. Some peculiarities for a two SQUID arrangement with 4 Josephson junctions are observed and simulated on the base of the RSJ model. >
4
dc SQUID magnetometers from single layers of YBa2Cu3O7−<i>x</i>
D. Koelle, A. H. Miklich, Frank Ludwig et al. · Applied Physics Letters · 1993 · 115 citations
A second-order SQUID gradiometer operating at 77 K
Y. Tavrin, Y Zhang, Werner Wolf et al. · Superconductor Science and Technology · 1994 · 69 citations · Full text