International Journal of Environmental & Analytical Chemistry · 1993 · 10 citations · 6 references
Natural SciencesChemistryEnvironmental SciencesAnalytical ChemistryElemental AnalysisInorganic ChemistryEnvironmental Analytical ChemistryHeavy Metal AnalysisHeavy Metal ContaminationInorganic Elemental AnalysisTrace Element AnalysisWater QualityAnalytical Detection Of Heavy MetalsGraphite Furnace Atomic Absorption SpectrometryGraphite Furnace Atomic Absorption SpectroscopyHeavy Metal Contamination AssessmentHeavy Metal DeterminationTrace Metal AnalysisWater Quality Elemental Analysis
Abstract Summary-Two methods for the determination of cadmium, copper and lead in EPA reference precipitation sample, corresponding to 1987, 1988, 1989 and 1990 BAPMoN Intercomparison Program, have been compared: graphite furnace atomic absorption spectroscopy (GFAAS) and anodic stripping voltametry (ASV). A relatively good agreement was found between the two techniques and with the reference values. The results for the analysis of Cl and F by spectrophotometry, Ca, Mg, Na, K and Zn by AAS, and Fe, Mn, Cu, Ni, Pb and Cd by GFAAS in the same sample, corresponding to 1988 and 1989, showed good agreement with the stated values. The results of Easter Island rain waters for the years 1990 and 1991 are presented.
6