Accuracy of the Unipolar Electrogram for Identification of the Site of Origin of Ventricular Activation

54

Citations

15

References

1997

Year

Abstract

While an R wave in the unipolar electrogram can be seen as close as 2 mm from the site of impulse origin, the absence of an R wave as an indicator of the site of impulse origin in the right ventricle is highly inaccurate. Therefore, the absence of an R wave in the unipolar electrogram is unlikely to be an adequate guide for identification of an effective target site for ablation of right ventricular tachycardia.