A 60-dB dynamic-range CMOS sixth-order 2.4-Hz low-pass filter for medical applications
224
Citations
15
References
2000
Year
Low-power ElectronicsElectrical EngineeringEngineeringMixed-signal Integrated CircuitAnalog DesignHz Low-pass FitterFilter DesignDigital FilterGrounded CapacitorsMicroelectronicsMedical ApplicationsDynamic Range
The design and implementation of a fully integrated complementary metal-oxide-semiconductor (CMOS) sixth-order 2.4 Hz low-pass fitter (LPF) for medical applications is presented. For the implementation of large-time constants both linearized operational transconductance amplifiers with reduced transconductance and impedance scalers schemes for grounded capacitors are employed. Experimental results for the filter have shown a dynamic range (DR) of 60 dB, while the harmonic distortion components are below -50 dB. The power consumption for the filter is below 10 /spl mu/W, the power supply is /spl plusmn/1.5 V, and the active area is 1 mm/sup 2/. The filter was fabricated in a double poly double metal 0.8 /spl mu/m CMOS process.