IEEE Transactions on Instrumentation and Measurement · 1997 · 53 citations · 3 references
Natural SciencesPhysicsGravitationGravityUncertainty QuantificationCalibrationGravity MeasurementsMeasurement ErrorMeasurement UncertaintyAlignment CorrectionMagnetic FluxMagnetic Flux DensityMisalignment CompensationMisalignment Effects
The effects of alignment uncertainties of the NIST watt balance with respect to local gravity and the magnetic flux density of the balance have been analyzed. Techniques for measuring all quantities relevant to misalignment have been developed. The components of the relative combined standard uncertainty of the measured value of the watt due to alignment uncertainties have been reduced to 20 nW/W, and potential improvements in the balance design have been identified which could ultimately lead to a reduction of that uncertainty to below 10 nW/W.
3
Methods for aligning the NIST watt-balance
Gerard N. Stenbakken, Reto Steiner, P. T. Olsen et al. · IEEE Transactions on Instrumentation and Measurement · 1996 · 13 citations · Full text