Journal of Bone and Joint Surgery - British Volume · 2007 · 55 citations · 57 references
MedicineOrthopaedic SurgeryOrthopedic SurgeryBiomechanicsOrthopedic BiomechanicsOrthopedicsArthroplasty BiomechanicsBone BiomechanicsBone MechanicsHip ArthroplastyHip BiomechanicsJoint MechanicsContact StressesHip Contact MechanicsHip Implant BiomechanicsHip Joint LoadingHip Joint MechanicsInitial Mechanical StabilityJoint Contact MechanicsTotal Hip Arthroplasty Biomechanics
Finite element analysis was used to examine the initial stability after hip resurfacing and the effect of the procedure on the contact mechanics at the articulating surfaces. Models were created with the components positioned anatomically and loaded physiologically through major muscle forces. Total micromovement of less than 10 mum was predicted for the press-fit acetabular components models, much below the 50 mum limit required to encourage osseointegration. Relatively high compressive acetabular and contact stresses were observed in these models. The press-fit procedure showed a moderate influence on the contact mechanics at the bearing surfaces, but produced marked deformation of the acetabular components. No edge contact was predicted for the acetabular components studied. It is concluded that the frictional compressive stresses generated by the 1 mm to 2 mm interference-fit acetabular components, together with the minimal micromovement, would provide adequate stability for the implant, at least in the immediate post-operative situation.