Recueil des Travaux Chimiques des Pays-Bas · 1960 · 69 citations · 2 references
Natural SciencesChemistryAnalytical ChemistryElectroanalytical MethodsElectrochemistryElectrolyte SolutionsPh AnalysisSolution ChemistryAcid-base BufferingInterfacial PhLiquid Junction PotentialsPh MeasurementPh Meter Calibration
Abstract The determination of the liquid‐junction potential at the interface of an oxalate and a succinate buffer solution of known pH* in water, methanol or a methanol‐water mixture and a saturated KCl solution in water at 25° is described. When the pH* of such a solution is measured by means of a pH meter standardized by an aqueous standard solution, this liquid‐junction potential causes an error of over 2 pH* when the methanol content of the solution is high.
2
Roger G. Bates · Journal of research of the National Bureau of Standards · 1947 · 14 citations · Full text