British Journal of Radiology · 1951 · 21 citations · 0 references
MedicineAnatomyRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyBone Quantity AssessmentImaging AssessmentRadiographic AssessmentSoft Tissue Mass ImagingBone Density EstimationBone MorphometryBone Volume MeasurementComputed Tomography-based MeasurementCortical Bone DensityExperimental SubjectRadiographic Bone MeasurementSpecular Reflection
The radiological measurement of bone density has been based upon mathematics unsupported by experiment. Experiments were, therefore, developed to prove the mathematical argument. These showed the importance of the mass of the soft tissue and suggested a method of standardising this mass effect. A wedge (Fig. 1) was cut from the cortex of the mid-shaft of an ox femur. Over a known region of this wedge, when the image of the wedge is on the straight line portion of the Hurter and Driffield curve of the film, D = kh = k′L, where D = specular density, k and k′ = constants, h = thickness of the wedge, and L = length of the wedge. The length of the wedge is thus a convenient measure of its thickness. The specular densities were measured or compared by means of a Weston Photographic Analyser, Model 877, using a field 1·0 mm. in diameter. A tray made of 0·45 cm. Perspex, 6 cm. deep, with an area of 14 cm. by 24·9 cm., was used to hold the experiment material. The phantom or experimental subject and the wedge were radiographed simultaneously on a single film. The specular density at a particular point was measured, and the thickness of the wedge producing the same density was found. This will be called the Equivalent Wedge Thickness (EWT) of the phantom or subject at that point.