A Device for Dried Blood Microsampling in Quantitative Bioanalysis: Overcoming The Issues Associated Blood Hematocrit

Neil Spooner, Philip Denniff, L. Michielsen, Ronald de Vries, Qin Ji, Mark E. Arnold, Karen Woods, Eric Woolf, Yang Xu, Valérie Boutet,

Bioanalysis · 2014 · 217 citations · 24 references

DOIFull text

Open access

Concepts

Abstract

AIMS: A cross-laboratory experiment has been performed on a novel dried blood sampler in order to investigate whether it overcomes issues associated with blood volume and hematocrit (HCT) that are observed when taking a subpunch from dried blood spot samples. MATERIALS & METHODS: An average blood volume of 10.6 μl was absorbed by the samplers across the different HCTs investigated (20-65%). RESULTS: No notable change of volume absorbed was noted across the HCT range. Furthermore, the variation in blood sample volumes across six different laboratories was within acceptable limits. CONCLUSION: The novel volumetric absorptive microsampling device has the potential to deliver the advantages of dried blood spot sampling while overcoming some of the issues associated with the technology.

References

24