Journal of Cardiovascular Electrophysiology · 2013 · 132 citations · 7 references
In patients without an indication for pacing, 7.4% would have been not suitable for a S-ICD according to the TMS. HCM, a heavy weight, a prolonged QRS duration and a R:T ratio <3 in the ECG lead with the largest T wave were independently associated with TMS failure. These data might alert physicians that selection of patients for a S-ICD should be considered with special caution in certain patient groups, because they may not satisfy ECG criteria for adequate sensing.
7
An Entirely Subcutaneous Implantable Cardioverter–Defibrillator
Gust H. Bardy, Warren Smith, Margaret Hood et al. · New England Journal of Medicine · 2010 · 762 citations · Full text