Publication | Closed Access
Modes of Implant Failure After Hip Resurfacing: Morphological and Wear Analysis of 267 Retrieval Specimens
260
Citations
26
References
2008
Year
Failures on the femoral side usually occur within the first nine months after surgery and appear to be most directly related to the implantation technique or patient selection. Later failures are observed mainly due to acetabular problems, either due to dramatically increased wear or poor cup anchorage. Improper cup anteversion may be similar to or more important than cup inclination in producing excessive wear.